Experimental study of multi-photon contamination on the measurement offluorescent decay time
- Received Date: 2008-03-13
- Accepted Date: 2008-05-28
- Available Online: 2008-11-05
Abstract:
In the measurement of fluorescent lifetime based on time correlation-single photon counting technique by means of TAC, due to the contamination of multi-photons a deviation of fluorescent lifetime measured from the expected value is experimentally studied. A correction function instead of a simple exponential function is used to fit the experiment data. The validation of the correction function is checked using the experimental data of several test samples: YAP, NaI(Tl) and LSO. The results show that the correction function well fits the data and the reasonable fluorescent lifetimes are obtained.