An MRPC for fast neutron detection

Get Citation
WANG Yi, CHENG Jian-Ping, LI Yuan-Jing, LUO Ming and ZHANG Guo-Guang. An MRPC for fast neutron detection[J]. Chinese Physics C, 2010, 34(1): 88-91. doi: 10.1088/1674-1137/34/1/016
WANG Yi, CHENG Jian-Ping, LI Yuan-Jing, LUO Ming and ZHANG Guo-Guang. An MRPC for fast neutron detection[J]. Chinese Physics C, 2010, 34(1): 88-91.  doi: 10.1088/1674-1137/34/1/016 shu
Milestone
Received: 2009-02-18
Revised: 2009-03-17
Article Metric

Article Views(2802)
PDF Downloads(789)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

An MRPC for fast neutron detection

    Corresponding author: WANG Yi,

Abstract: 

The possibility to detect fast neutrons with a multi-gap resistive plate chamber (MRPC) has been investigated. To detect fast neutrons, a thin polyethylene layer is coated on the surface of electrode glass as a fast neutron converter. The MRPC detects the charged particles generated by neutrons via the (n,p) reaction on hydrogen. A prototype detector has been developed and tested on fast neutron sources in order to evaluate its performance: good agreement between experimental results and simulation has been achieved. A detailed
description of the detector and the experimental test results are presented.