Fluence rate linear response limit of a scintillators to γ-rays

Get Citation
GUAN Xing-Yin, SONG Zhao-Hui and HAN He-Tong. Fluence rate linear response limit of a scintillators to γ-rays[J]. Chinese Physics C, 2011, 35(7): 656-659. doi: 10.1088/1674-1137/35/7/010
GUAN Xing-Yin, SONG Zhao-Hui and HAN He-Tong. Fluence rate linear response limit of a scintillators to γ-rays[J]. Chinese Physics C, 2011, 35(7): 656-659.  doi: 10.1088/1674-1137/35/7/010 shu
Milestone
Received: 2010-09-25
Revised: 2010-11-08
Article Metric

Article Views(2744)
PDF Downloads(386)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Fluence rate linear response limit of a scintillators to γ-rays

Abstract: Scintillation detectors based on LSO, CeF and PbWO are the main candidates for measuring γ-rays in a mixed γ/n pulsed radiation field with high intensity. An experiment using the Lissajous figure method to study the high fluence rate response behavior of three kinds of commonly used scintillators is introduced in this paper. The result shows that the fluence rate linear response limit of LSO and CeF is 1.9×1019 and 2.1×1018MeV/(cm2·s), respectively, and the PbWO scintillator still maintains linear response when the fluence rate of γ-ray is up to 2.0 × 1020 MeV/(cm2·s).