Trapezoidal pulse shaping for pile-up pulse identification in X-ray spectrometry

Get Citation
ZHOU Jian-Bin, LIU Yi, HONG Xu, ZHOU Jin, MA Ying-Jie, WANG Min, HU Yun-Chuan, CHEN Bao and YUE Ai-Zhong. Trapezoidal pulse shaping for pile-up pulse identification in X-ray spectrometry[J]. Chinese Physics C, 2015, 39(6): 068201. doi: 10.1088/1674-1137/39/6/068201
ZHOU Jian-Bin, LIU Yi, HONG Xu, ZHOU Jin, MA Ying-Jie, WANG Min, HU Yun-Chuan, CHEN Bao and YUE Ai-Zhong. Trapezoidal pulse shaping for pile-up pulse identification in X-ray spectrometry[J]. Chinese Physics C, 2015, 39(6): 068201.  doi: 10.1088/1674-1137/39/6/068201 shu
Milestone
Received: 2014-09-28
Revised: 2014-12-19
Article Metric

Article Views(1890)
PDF Downloads(131)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Trapezoidal pulse shaping for pile-up pulse identification in X-ray spectrometry

    Corresponding author: LIU Yi,

Abstract: Energy resolution is affected by the intrinsic energy resolution of the detector, ballistic deficit, pile-up pulses and noise. Pile-up pulses become the dominant factor that degrades energy resolution after the system is established, so pile-up rejection is often applied to obtain good energy resolution by discarding pulses that are expected to be contaminated by pile-up. However, pile-up rejection can reduce count rates and thus lower the measurement precision. In order to improve count rates and maintain energy resolution, a new method of pile-up pulse identification based on trapezoidal pulse shaping is presented. Combined with pulse width discrimination, this method is implemented by recording pulses that are not seriously piled up. Some experimental tests with a Cu-Pb alloy sample are carried out to verify the performance of this method in X-ray spectrometry. The results show that the method can significantly improve count rates without degrading energy resolution.

    HTML

Reference (1)

目录

/

DownLoad:  Full-Size Img  PowerPoint
Return
Return