Experimental Study on the Method of Deadtime Correction For HPGe Spectrometer System

  • A practical method of deadtime correction for HPGe γ spectrometer system was given.The relation between the percent deadtime and the correction of counting rate losses for HPGe γ spectrometer system was found by using a strong γ source.Therefore, the activity measurement can be implemented at relatively high count rate,with this γ spectrometer.
  • 加载中
  • [1] . XIANG Si-Cang, WANG Yong-Chang, YUAN Jun-Qian et al. Nuclear Techniques, 1991, 14(6): 355—358(in Chinese)(项斯苍,王永昌,袁俊谦等.核技术,1991, 14(6): 355-358)2. HE Fu-Qing, LONG Xian-Guan, LIU Man-Tian et al. Nu-clear Electronics Detection Technology, 1991, 11(3):177—179(in Chinese)(何福庆,龙先灌,刘慢天等.核电子学与探测技术,1991,11(3): 177-179)3. HE Fu-Qing, LONG Xian-Guan, PENG Xiu-Feng et al.Nuclear Electronics Detection Technology, 1996, 16(2):142—143(in Chinese)(何福庆,龙先灌,彭秀峰等.核电子学与探测技术,1996,16(2):142-1434. GU Dang-Chang. Method Technology of Activity Mea-surements of Radionuclide. Beijing: Science Press,1994,257—258(in Chinese)(古当长.放射性核素活度测量的方法和技术.北京:科学出版社,1994, 257-258)5. Firestone R B, Shirley V S. Table of Isotopes. New York,19966. Fudan University, Qinghua University, Beijing University.Nuclear Experimental Method(The Third Edition). Bei-jing: Atomic Energy Press, 1997, 6—13(in Chinese)(复旦大学,清华大学,北京大学.原子核物理实验方法(第3版).北京:原子能出版社,1997, 6-13)
  • 加载中

Get Citation
ZHOU Feng-Qun, YANG Jing-Kang, TUO Fei, YI Yan-Ling and KONG Xiang-Zhong. Experimental Study on the Method of Deadtime Correction For HPGe Spectrometer System[J]. Chinese Physics C, 2005, 29(12): 1175-1178.
ZHOU Feng-Qun, YANG Jing-Kang, TUO Fei, YI Yan-Ling and KONG Xiang-Zhong. Experimental Study on the Method of Deadtime Correction For HPGe Spectrometer System[J]. Chinese Physics C, 2005, 29(12): 1175-1178. shu
Milestone
Received: 2005-03-11
Revised: 1900-01-01
Article Metric

Article Views(4310)
PDF Downloads(767)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Experimental Study on the Method of Deadtime Correction For HPGe Spectrometer System

    Corresponding author: ZHOU Feng-Qun,
  • Department of Modern Physics,Lanzhou University,Lanzhou 730000,China2 Department of Physics,Pingdingshan University,Pingdingshan 467000,China

Abstract: A practical method of deadtime correction for HPGe γ spectrometer system was given.The relation between the percent deadtime and the correction of counting rate losses for HPGe γ spectrometer system was found by using a strong γ source.Therefore, the activity measurement can be implemented at relatively high count rate,with this γ spectrometer.