Monte Carlo Simulations for Optimum Design of Composite Detectors for High Energy γ Ray

  • The efficiencies of clover and cluster composite detectors using NaI and BGO crystals as the media for detection of high energy γ ray are systematically simulated with Monte Carlo method. It is shown that for the same geometry of detection media concerned the efficiency of the composite BGO detector is much higher than that of the composite NaI detector. Therefore NaI crystal is not a suitable medium of composite detectors for high energy γ ray duo to low efficiency, Doppler broadening and distortion to γ spectrum in comparison with BGO crystal. The composite BGO detectors have many advantages such as large photopeak efficiency, small Doppler effect and regular γ spectrum. As to the clover and cluster composite detectors consisting of the cylinders of BGO crystal with original size 76×127, the intrinsic photopeak efficiencies are over 40% and the enhanced factor of absolute efficiencies is as high as 2.4 and 2.7, respectively, for 22 MeV γ ray.
  • 加载中
  • [1] . Duchene G ,Beck F A ,Twin P J et al.Nucl.Instr.Meth.,1999,A432:90—110 2. Million B ,Bracco A ,Camera F et al.Nucl.Instr.Meth.,2000,A452:422—430 3. Shepherd S L ,Nolan P J,Cullen D M et al.Nucl.Instr.Meth.,1999,A434:73—3864. Maj A ,Gaardhoje J J,Atac A et al.Nucl.Phys.,1994,A571:185—220 5. Bruce A M ,Gaardhoje J,Herskind B et al.Phys.Lett.,1988,B215:237—2416. WEN WanXin,JIN GenMing.High Ener.Phys.and Nucl.Phys.,2002,26:1178(in Chinese)(文万信,靳根明.高能物理与核物理,2002,26:1178)7. Csatlos M ,Krasznahorkay A ,Bacelar J C S et al.Nucl.Instr.Meth.,1999,A421:272—2778. Brun R ,Carminati F .GEANT Detector Description and Simulation Tool,CERN Program Library,Long Writeup W5013ed.,CERN ,Geneva,19949. WEN WanXin,JIN GenMing,WANG SuFang.Nucl.Phys.Rev.,1999,16:237—242(in Chinese)(文万信,靳根明,王素芳.原子核物理评论,1999,16:237—242)
  • 加载中

Get Citation
WEN Wan-Xin and JIN Gen-Ming. Monte Carlo Simulations for Optimum Design of Composite Detectors for High Energy γ Ray[J]. Chinese Physics C, 2002, 26(12): 1291-1296.
WEN Wan-Xin and JIN Gen-Ming. Monte Carlo Simulations for Optimum Design of Composite Detectors for High Energy γ Ray[J]. Chinese Physics C, 2002, 26(12): 1291-1296. shu
Milestone
Received: 2001-12-29
Revised: 1900-01-01
Article Metric

Article Views(3147)
PDF Downloads(709)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Monte Carlo Simulations for Optimum Design of Composite Detectors for High Energy γ Ray

    Corresponding author: WEN Wan-Xin,
  • Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China2 Nuclear Medicine School, Soochow University, Suzhou 215007, China

Abstract: The efficiencies of clover and cluster composite detectors using NaI and BGO crystals as the media for detection of high energy γ ray are systematically simulated with Monte Carlo method. It is shown that for the same geometry of detection media concerned the efficiency of the composite BGO detector is much higher than that of the composite NaI detector. Therefore NaI crystal is not a suitable medium of composite detectors for high energy γ ray duo to low efficiency, Doppler broadening and distortion to γ spectrum in comparison with BGO crystal. The composite BGO detectors have many advantages such as large photopeak efficiency, small Doppler effect and regular γ spectrum. As to the clover and cluster composite detectors consisting of the cylinders of BGO crystal with original size 76×127, the intrinsic photopeak efficiencies are over 40% and the enhanced factor of absolute efficiencies is as high as 2.4 and 2.7, respectively, for 22 MeV γ ray.