THE TIME-METHOD FOR THE DETERMINATION OF THE POSITION IN THE FOUR OUTPUTS SCINTILLATION HODOSCOPE
- Received Date: 1900-01-01
- Accepted Date: 1900-01-01
- Available Online: 1983-08-05
Abstract: Using the relation of time of four outputs in the scintillation hodoscope (50cm×50cm×0.5cm), the position of incident particles is determined.
The time resolution σt=140—360 ps(correspondingly the position resolution σL=2—5 cm) has been obtained.